@phdthesis{digilib65647, month = {May}, title = {VALIDASI UKURAN LOLOSAN 99Mo DARI GENERATOR 99Mo/99mTc NON-FISI UNTUK MENGHASILKAN RADIONUKLIDA MEDIS 99mTc}, school = {UIN SUNAN KALIJAGA YOGYAKARTA}, author = {NIM.: 20106020034 Nur Zulfa Ramadhanti}, year = {2024}, note = {Pembimbing: Siska Febriana, S.Si., M.Sc. dan Dr. Nita Handayani, S.Si., M.Si}, keywords = {Generator 99Mo/99mTc, Molibdenum-99, Spektrometer Gamma, Teknesium-99m}, url = {https://digilib.uin-suka.ac.id/id/eprint/65647/}, abstract = {The medical radionuclide 99mTc is the most widely used type of radionuclide in the field of nuclear medicine. 99mTc is the decay product of 99Mo. One of the quality control parameters of 99mTc is the 99Mo breakthrough limit. It is important to know the 99Mo breakthrough so that the quality of 99mTc produced can be suitable and optimal for diagnosis in nuclear medicine. This study aims to validate the 99Mo breaktroughs activity from a non-fission 99Mo /99mTc generator whether they meet the established international limit standard. This reseach used a non-fission 99Mo /99mTc generator produced in multipurpose reactor G.A. Siwabessy, Indonesia and a mixed liquid standard source of 133Ba and 152Eu. The measurement of 99Mo breakthrough using gamma spectrometer with High Purity Germanium (HPGe) detector. The results obtained showed that the 99Mo elution in the first elution had the highest elution of 0,233510238 uCi 99Mo /mCi 99mTc, which exceeded the set limit of 0.15 uCi 99Mo /mCi 99mTc. While in the second, third, fourth, and fifth elution have the 99Mo breakthrough, but these breakthrough did not exceed the limit and met one of the requirements of the 99mTc medical radionuclide parameter.} }